• +86-13271526781
  • thermo@gwdl.com

GWDLTHERM 1100-degree-small-vertical-bular-sistance-nace

GWDLTHERM 1100°C vertical tube furnace features dual-zone 800mm heating, 108mm stainless steel tube, KF flange sealing, vacuum pump, inert gas inlets and 7-inch touch screen control. ±1°C accuracy.

GWDLTHERM 1100-degree-small-vertical-bular-sistance-nace

GWDLTHERM 1100-degree-small-vertical-bular-sistance-nace

Product Name:1100°C Vertical Tube Furnace
Maximum Temperature:1100°C
Total Heating Zone Length:800 mm
Furnace Tube Total Length:1.3 m
Furnace Tube Outer Diameter:108 mm
Furnace Tube Sealing:304 Stainless Steel KF Flange

1100°C Vertical Tube Furnace — Technical Specifications

ItemSpecification
Maximum temperature1100°C
Voltage (AC)380 V
Long-term working temperature1100°C
Temperature control accuracy±1°C
Furnace tube materialStainless steel tube, outer diameter 108 mm
Furnace tube sealing304 stainless steel KF flange; fixing devices added at both ends of the tube to prevent the tube from falling over
Heating zone size2 temperature zones, each 400 mm; total heating zone 800 mm; total tube length 1.3 m
Gas inletTwo inlets, flow meter 0–10 L/min
Vacuum pumpMechanical vacuum pump, vacuum degree −0.1 MPa
Temperature field uniformity inside the furnace±3°C
Temperature sensing element and rangeK-type thermocouple, range 0–1320°C
Furnace bodyCNC-machined; polished, ground, acid-washed, phosphated, powder-coated and high-temperature baked; two-color design, novel and attractive appearance; oxidation-resistant, acid/alkali-resistant, corrosion-resistant, high-temperature-resistant and easy to clean
Furnace body structureInternationally advanced air-cooled double-layer furnace shell structure; effective air-cooled guide baffle enables overall cold-air circulation in the shell; finally cools the conductive plates of heating elements before discharging from the furnace body, avoiding high-temperature oxidation of the conductive plates and ensuring a good working environment
Furnace door openingSide opening
Refractory materialHigh-purity alumina fiber, vacuum-formed fiber lightweight board material for the furnace lining; light-weight hollow-sphere alumina boards used at positions easily hit during loading/unloading (furnace mouth, furnace bottom); high service temperature, low heat storage, resistance to rapid heating/cooling, no cracking, no slagging, good insulation performance (energy-saving effect over 80% compared with old-style electric furnaces)
Insulation materialThree-layer insulation: alumina fiber cotton, alumina fiber board, alumina polycrystalline fiber board; energy-saving effect over 80% compared with old-style electric furnaces
ControlClosed-loop technology, SCR module trigger control, phase-shift triggering or zero-crossing triggering; output voltage, current or power continuously adjustable; constant voltage, constant current or constant power characteristics; current loop as inner loop and voltage loop as outer loop; when load is suddenly applied or load current exceeds the limit value, the output current of the voltage regulator is limited within the rated current range to ensure normal operation of the output and voltage regulator; meanwhile the voltage loop also participates in regulation, so that the output current of the voltage regulator is limited within the rated current range and the output current and voltage are maintained constant under the premise of sufficient regulation margin; thereby protecting heating elements from impact of excessive current and voltage and achieving safe, reliable control effect and control accuracy
Heating elementResistance wire
Gases allowedNitrogen, argon, hydrogen and other inert gases
Heating element positionInstalled around the furnace chamber
Refractory materialHigh-purity alumina fiber board
Cooling structureDouble-layer furnace shell, air cooling
Furnace body temperature≤45°C
Warranty scope and periodFree warranty for the electric furnace for one year; heating elements and furnace tube not covered (heating elements naturally damaged within three months will be replaced free of charge)
Additional configuration7-inch true-color touch screen
Random accessoriesOne manual, one certificate of conformity, one pair of high-temperature gloves

 

GWL Touch Screen Control Introduction

 

ItemParameter
Size7-inch TFT true color
Resolution800 × 480
BacklightLED
Power5 W
Weight1 kg
LanguageChinese/English, freely switchable
Full-screen display and operationInstrument screen, bar graph, historical trend, data report, alarm information, data export, process flow, system management, etc.
Touch screen operationTouch operation and mouse operation can be used arbitrarily

Touch Screen Control Functions

 

 

ItemSpecification
Controlled objectsTemperature, pressure, flow, liquid level, etc.
Number of furnaces controlled1 unit
Temperature controlTouch screen + high-precision integrated module
Start heatingTouch operation
Pause heatingTouch operation
Stop heatingTouch operation
Temperature curve programmingTouch operation, 30 segments per curve
Number of stored temperature curvesUnlimited (each curve can be named in Chinese/English)
Real-time status displayRunning curve name, running segment number, segment time, segment running time, digital temperature, real-time curve, power output percentage
Start from selected segment (cross-segment start)Touch operation
Curve rangeAdjustable
Historical curve (graphic/text)Approximately 20 months of storage
Data report (EXCEL)Approximately 20 months of storage; multi-point simultaneous display (USB flash drive can be inserted for expansion)
Historical curve and report record (save) intervalAdjustable from 1 s to 3600 s; multi-point simultaneous display (USB flash drive can be inserted for expansion)
Alarm indicationColor change (red)
Alarm information languageChinese display (alarm time and event described in Chinese)
Data export interfaceUSB
Printer interfaceParallel port
Touch screen operation protectionPassword setting (others cannot operate without password)
Communication portRS485
PasswordPasswords can be set for all touch screen operations, preventing leakage of parameters and processes and enabling comprehensive management
Connect with Us Today
Phone Phone:+86-13271526781
Email Email:thermo@gwdl.com