Furnace size | ≥400*400*400 mm |
highest temperature | ≥1400℃ |
Temperature control accuracy | Better than or equal to ±1 ℃ |
Sample platform | Equipped with electric rotation function, sample platform diameter ≥360 mm |
Temperature uniformity | Better than or equal to ±3℃ (within the effective furnace volume) |
cooling rate | Temperature drop rate adjustable from 0.1 to 5℃/min within the temperature range of 500℃-1200℃. |
Temperature controller | The standard configuration includes a digital controller that can be programmed based on heating and cooling rates and time. It features automatic temperature control and time management, and the digital control system can accurately set and display the temperature. |
Protection function | It features an automatic heating stop function when the door is opened, an alarm system for abnormal temperature and equipment malfunction, and automatic power-off protection against leakage. |
High temperature protection | The machine's outer surface temperature is <50℃ at the highest operating temperature. |
Process security system | Passwords can be set to protect process parameters. |