• +86-13271526781
  • thermo@gwdl.com

1000°C Pneumatic Top Opening Box Type Muffle Furnace for Laboratory Heat Treatment

This laboratory muffle furnace is specially designed for research laboratories, universities, and testing institutions. It provides precise temperature control, stable performance, and long-term reliability for various laboratory heat treatment applications

1000°C Pneumatic Top Opening Box Type Muffle Furnace for Laboratory Heat Treatment

1000°C Pneumatic Top Opening Box Type Muffle Furnace for Laboratory Heat Treatment

Maximum working temperature:1000°C / 1200°C
Long-term operating temperature:700-900 degrees
Temperature control range:100-1000 degrees
Furnace size:500 × 500 × 500mm
Temperature control accuracy:±1 degree
Temperature sensing element:K -type thermocouple

This 1000°C pneumatic top-opening box type electric muffle furnace is designed for laboratory heat treatment applications requiring convenient operation and precise temperature control. The pneumatic lifting structure allows easy loading and unloading of samples, improving laboratory efficiency and safety.

Equipped with a programmable PID temperature controller, the furnace ensures accurate temperature regulation and excellent uniformity throughout the chamber. High-performance insulation materials reduce heat loss and energy consumption while maintaining stable long-term operation.

Typical applications include material research, ceramic sintering, metal annealing, chemical analysis, and laboratory testing in universities and research institutes.

Warranty & Service:
• 1-year warranty
• Lifetime technical support and after-sales service
• Reliable performance for continuous laboratory use
 

Maximum operating temperature

1000 degrees

Long-term operating temperature

700-900 degrees

 

Temperature control range

 

100-1000 degrees

 

Furnace size

500 × 500 × 500mm

Temperature control accuracy

±1 degree

Temperature sensing element

K -type thermocouple

 

Heating element material

 

High-temperature alloy resistance wire

Heating element installation location

 

Four-sided heating

 

Fully automatic intake and exhaust ports

A protective atmosphere can be introduced; the electric exhaust port and electric intake port can be linked to the temperature rise curve segment selection and automatically open and close at any temperature range.
heating rate

Minimum 10-30°C/ minute

 

 

Furnace body structure and materials

The electric furnace body adopts an advanced air-cooled double-layer carbon steel furnace body structure, with effective air cooling guidance. The baffles allow for the circulation of cold air throughout the furnace shell, which ultimately cools the conductive plates of the heating elements before being discharged from the furnace. This design avoids high-temperature oxidation of the conductive sheets in the heating element, ensuring a good working environment.

territory.

 

 

Stainless steel fan

 

Stainless steel fans assist chimneys for forced ventilation

 

Observation windowMinimum size 30 x 35 mm, Material / Quartz Glass
Door opening method

Electric top-opening door

 

Refractory insulation materials

The furnace lining is made of vacuum-formed high-purity alumina lightweight material, which has high operating temperature, low heat storage, resistance to rapid heating and cooling, no cracking, no slag shedding, and good heat insulation performance.
Furnace shell temperatureThe casing temperature is less than 40 degrees Celsius.

 

 

Security Protection

An integrated modular control unit is adopted, ensuring accurate control precision, and a dual-loop control system is designed. It features dual-circuit protection, including overshoot, over-adjustment, under-adjustment, thermocouple breakage, phase loss, overvoltage, overcurrent, overtemperature, current feedback, and soft start protection.

 

Power off when door is opened

The furnace body is designed with a function that opens the furnace door and disconnects the main heating circuit when the temperature is high during operation, greatly protecting against the risk of electric shock when handling materials at high temperatures.

 

 

 

 

 

Safe temperature control

Thyristor module triggering control using closed-loop technology, phase-shift triggering control, or zero-crossing triggering. The output voltage, current, or power is continuously adjustable, and it features constant voltage, constant current, or... Constant power characteristics; the current loop is the inner loop, and the voltage loop is the outer loop. This is important when a sudden load or negative voltage is applied. When the load current exceeds the current limit, the voltage regulator's output current is limited to the rated current range. Internally, this ensures the output and voltage regulator function properly; simultaneously, the voltage loop also participates in regulation, making the regulator... The output current of the transformer is limited to the rated current range, with sufficient adjustment margin. Under the premise of maintaining constant output current and voltage, thereby protecting the heating element from overheating. It can withstand the impact of excessive current and voltage to achieve a safe and reliable control effect and control accuracy.

 

 

 

 

Temperature profile setting

Employs an intelligent temperature controller, equipped with standard PID and AI-controlled APID. or It features multiple adjustment modes such as MPT , and has self-tuning and self-learning functions, with no overshoot or undershoot. Excellent control characteristics, with 50 The segmented program control function can achieve arbitrary slopes.

features temperature control with programmable/operable commands such as jump (cycle ) , run, pause, and stop , and allows modification of the program at any time during its operation. It employs an artificial intelligence adjustment algorithm with curve fitting capabilities to achieve smooth and even curve control.

 

 

Number of segments in the heating curve

50 The segmented program control function allows input of settings: a curve is... 50 Section, two melodies

Wire 28 Segment / strip, three curves 15 Segment / segment, five curves 9 Segment /line; Multiple curves can be entered simultaneously and can be called up arbitrarily during use.

Panel buttonsThe two buttons are: main power button/knob and heating chamber on button/knob.
Random accessoriesOne crucible tong, two pairs of high-temperature gloves, and one furnace bottom pad.

 

 

III. Warranty Scope and Period

The electric furnace comes with a one-year free warranty, but the furnace bottom plate and heating elements are not covered by the warranty (they will be replaced free of charge if they fail naturally within three months).

IV. Packing List

High-temperature gloves 2 Secondary; Corundum furnace bottom plate 1 Block; crucible tongs 1 The host manual 1 Copies; Warranty Card 1 Shares; personnel

One operating and usage training copy; one delivery note and one equipment acceptance form.

V. Precautions

1. To avoid affecting the lifespan of the electric furnace, it is recommended that the maximum heating and cooling rates be 1-15℃/min . (Rapid heating at high temperatures will shorten the lifespan of the heating element.)
2. This box furnace does not use a vacuum sealing structure, so flammable and explosive gases must not be introduced.
3 . After a period of use, minor cracks may appear in the furnace chamber of this box furnace. This is normal and will not affect its use. It can be repaired with an alumina coating.
4 . It is not recommended to introduce corrosive gases. If a highly corrosive gas must be introduced, such as... S, Na Please inform us in advance so we can perform special treatment on the furnace.
5. High-temperature solution must not leak onto the furnace bottom. To prevent this, a pad or alumina powder can be used for isolation.
6. The instrument should be placed in a well-ventilated, dry place.

VI. Shipping Details

1. The electric furnace is packaged in three layers: first wrapped in foam paper, then wrapped in plastic film, and finally packed in a wooden crate.
2. Free domestic door-to-door delivery (free delivery within city limits)
3. We will be responsible for any damage that occurs during the transportation of the electric furnace.
4. Logistics methods: transportation by road, rail, ship (for foreign trade export ) , and air (for foreign trade export). For nearby locations, our company will arrange dedicated vehicles for transportation (packaging is wooden pallets and cardboard boxes).

 

 

 

Connect with Us Today
Phone Phone:+86-13271526781
Email Email:thermo@gwdl.com